Paired hookwire electrode

Paired hookwire

 

 

The paired hookwire electrode consists of two 40cm PTFE insulated stainless steel wire AISI 302 (0.08mm d, 40G) threaded through a hypodermic needle. Ends of wires positioned so 2mm of one wire and 5mm of the second wire exit at end of needle.The first wire is stripped 2mm, while the second wire is insulated 3mm and stripped 2mm.The protruding end are bent at 180 degree from the needle. The other end of each wire is stripped 10mm.

 

TypeLeadQty.Item number
PHW - 100mm

(0,60 mm d,23G)
0,4 m cable

(0,08 mm d,40 G)
5/pkg.003-400161-5
PHW -50mm

(0,70mm d,(22G)
0,4 m cable

(0,08 mm d,40 G)
5/pkg.003-400162-5
PHW -30mm

(0,60mm d,(23G)
0,4 m cable

(0,08 mm d,40 G)
5/pkg.003-400163-5
 
FOR SINGLE PATIENT USE ONLY!
AVAILABLE BOTH STERILE OR READY FOR EtO STERILIZATION!